Model 4200A-SCS Parameter Analyzer Reference Manual

smeaszRT

This command makes and returns impedance measurements for a voltage or frequency sweep in real time.
Usage
int smeaszRT(int instr_id, long model, long speed, double *result1, char *colname1,
double *result2, char *colname2);
instr_id
model
speed
result1
colname1
result2
colname2
Details
This command is similar to the smeasz command; both commands return the measured impedance
readings for a sweep returned in arrays. However, the readings from smeaszRT are posted to the
Clarius Analyze sheet and graph in real time. Two measurement results appear in the sheet and
graph after each step of the sweep is executed.
Note that the values are only available in real-time if Clarius is running. Otherwise, they are stored in
an array.
The colname1 and colname2 parameters specify names for data sheet columns in Clarius.
Measurement model parameter values
Measurement model
ZTH
RjX
CpGp
CsRs
CpD
CsD
Example
smeaszRT(CVU1, 2, KI_NORMAL, result1, "result1", result2, "result2");
This command posts the results into the Clarius data sheet under columns named result1_arr and
result2_arr.
Also see
smeasz
(on page 13-189)
4200A-901-01 Rev. C / February 2017
The instrument identification code of the 4210-CVU: CVU1
Measure model (see "Measurement model parameter values" table in Details)
Speed settings:
KI_CVU_SPEED_FAST: Fast measurements (higher noise)
KI_CVU_SPEED_NORMAL: Selects a balance between speed and low noise
KI_CVU_SPEED_QUIET: Low-noise measurements
KI_CVU_SPEED_CUSTOM: Selects custom settings; the delay factor, filter
factor, and aperture are set using the setmode command
Array of the first result of the selected measure model
Column name to pass into Clarius for result1 array (character string)
Array of the second result of the selected measure model
Column name to pass into Clarius for result2 array (character string)
Impedance (Z) and phase (θ in degrees)
Resistance and reactance
Parallel capacitance and conductance
Series capacitance and resistance
Parallel capacitance and dissipation factor
Series capacitance and dissipation factor
Section 13: LPT library function reference
Parameter value
KI_CVU_TYPE_ZTH
or
KI_CVU_TYPE_RJX
or
or
KI_CVU_TYPE_CPGP
KI_CVU_TYPE_CSRS
or
KI_CVU_TYPE_CPD
or
KI_CVU_TYPE_CSD
or
0
1
2
3
4
5
13-191