Section 6: Clarius
Status codes
The 16 basic codes used for 4210-CVU measurement status are listed in table below. Each code is
represented as a 32-bit hexadecimal value (0x).
CVU measurement status codes (CVU1S)
#
Code
1
000000mr
2
8xxx00mr
3
01xx00mr
4
02xx00mr
5
03xx00mr
6
08xx00mr
7
09xx00mr
8
0Axx00mr
9
0Bxx00mr
10
xx0100mr
11
xx0200mr
12
xx0300mr
13
xx0800mr
14
xx0900mr
15
xx0A00mr
16
xx0B00mr
As shown in this table, the mr value is the last two digits of each code.
mr value
Measurement status notes
Whenever a fault occurs, run the Confidence Check utility before performing any other
troubleshooting actions (see
Measurement timeout: Indicates that the measurement was not received after a set time (total
aperture). This timeout error may indicate that there is an issue with the 4210-CVU card. Try resetting
the hardware and running the project test again. If this error reoccurs, contact Keithley Instruments.
To reset the hardware:
1. Select Start.
2. Type resethw.
3. Select the instruments that need reset.
4. Click Reset.
I measurement overflow: Try a higher I measure range (or Auto) or try a lower AC drive voltage.
V measurement overflow: Try a lower DC bias voltage.
ABB not locked: Auto Balance Bridge was not locked when the measurement was made. The
readings and calculation results may not be accurate.
6-192
Description
No faults
Measurement timeout occurred
CVH1 current measurement overflow
CVH1 voltage measurement overflow
CVH1 I and V measurement overflow
CVH1 ABB not locked
CVH1 ABB not locked, I measurement overflow
CVH1 ABB not locked, V measurement overflow
CVH1 ABB not locked, I and V measurement overflow
CVL1 I measurement overflow
CVL1 V measurement overflow
CVL1 I and V measurement overflow
CVL1 ABB not locked
CVL1 ABB not locked, I measurement overflow
CVL1 ABB not locked, V measurement overflow
CVL1 ABB not locked, I and V measurement overflow
00
Lowest range (1 µA) used for the impedance measurement
01
Middle range (30 µA) used for the impedance measurement
02
Highest range (1 mA) used for the impedance measurement
Confidence Check
Model 4200A-SCS Parameter Analyzer Reference Manual
(on page 4-19) for details).
4200A-901-01 Rev. C / February 2017