Clarius Project Example For Probe Sites; Cm500; Use Kcon To Add A Prober - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Appendix K: Using a Signatone CM500 Prober
7. In the File menu bar, select the Save setup As command to save the file to a hard disk. You can
load this setup next time without going through all of the procedures again.
8. The Prober is now ready to accept a remote command from the 4200A-SCS.

Clarius project example for probe sites

The following is a step-by-step procedure to configure the Clarius project to execute testing and
automatic wafer stepping to all programmed sites successfully. When the CM500 prober is connected
to the 4200A-SCS by GPIB interface, the 4200A-SCS is the GPIB master controller and the CM500 is
always in listening mode. The 4200A-SCS will send control commands to the CM500 to move the
prober to next site during the automatic testing. The interface commands are PrInit, PrChuck,
PrMovNxt, and PrSSMovNxt. You will need to add these commands into the Clarius project.

CM500

On the probe station computer, complete the

Use KCon to add a prober

You use KCon on the 4200A-SCS to add the prober to the configuration.
On the 4200A-SCS, use KCon to add the prober to the configuration:
1. Open KCon.
2. At the bottom of the Configuration Navigator, select Add External Instrument. The Add External
Instrument dialog box is displayed.
K-12
Model 4200A-SCS Parameter Analyzer Reference Manual
Figure 818: CM500 Prober Enable Subsite
Probe station configuration
Figure 819: Add a prober in KCon
(on page K-1) section.
4200A-901-01 Rev. C / February 2017

Advertisement

Table of Contents
loading

Table of Contents