Testing Flash Memory - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 6: Clarius

Testing flash memory

Clarius includes several projects that you can use to test floating gate transistors (NOR, NAND) and
other types of nonvolatile memory.
To use the flash memory tests, you will need:
Two pulse cards (four pulse channels).
At least two SMUs. If your system does not include switching, it is best to have four SMUs to
match the number of pulse channels to connect to a three- or four-terminal device. Either 4200-
SMUs or 4210-SMUs with SMU preamplifiers removed can be used.
Interconnecting cables and adapters, shown in the table below.
8 in. lb. torque wrench for tightening the SMA connections.
Recommended interconnect parts for flash memory testing
Qty.
Description
6
SMA tee, female – male – female
4
LEMO triaxial to SMA adapter
4
3-slot male triaxial to female BNC adapter
4
SMA male to BNC female adapter
2
10.8 cm (4.25 in.) white SMA cables
4
20.3 cm (8 in.) white SMA cables
4
2 m (6.6 ft) white SMA cables
4
1.5 m (5 ft) BNC cable
This configuration permits independent source and measure for each terminal in a typical 4-terminal
floating gate transistor.
6-336
Figure 392: Recommended parts for flash memory testing
Model 4200A-SCS Parameter Analyzer Reference Manual
Comment
Trigger, combine SMU and pulse channels
Adapt SMU Force output to SMA for signal
interconnect
Convert BNC cabling to triaxial for prober
or switch matrix connection
Adapt Tee to BNC for cabling from
instrument to probe manipulators
Interconnect for triggering
Interconnect between pulse card and SMU
signals
Connect to probe manipulators
4200A-901-01 Rev. C / February 2017
Keithley Model
Number
CS-1391
CS-1390
7078-TRX-BNC
CS-1252
CA-451A
CA-452A
CA-404B
7051-5