Model 4200A-SCS Parameter Analyzer Reference Manual
ki82ulib user library
The user modules in the ki82ulib user library control the Model 82 C-V System. They perform
simultaneous C-V, C-t, and Q/t measurements and cable compensation. The next table lists the user
modules. It also provides the name of tests and actions in Clarius that are based on these user
modules.
ki82ulib user modules
User module
Abortmodule82
29)
CableCompensate82
page D-30)
CTsweep82
DisplayCableCompCaps82
(on page D-36)
LoadCableCorrectionCon
stants82
QTsweep82
SaveCableCompCaps82
page D-41)
SIMCVsweep82
44)
Abortmodule82
The Abortmodule82() function puts the three System 82 instruments into a known state when a test is aborted.
This function is used by other library modules in the atexit() function.
Usage
atexit(Abortcleanup);
4200A-901-01 Rev. C / February 2017
Test and action
names
(on page D-
n/a
(on
cable-compensate
cablecomp
(on page D-33)
ctsweep
display-cap-file
n/a
(on page D-38)
qtsweep
(on
save-cap-file
savecablecompfile
(on page D-
system82-cvsweep
cvsweep
Appendix D: Using a Model 82 C-V System
Description
Puts the three System 82 instruments into a known
state when a test is aborted. This function is used by
other library modules in the atexit() function.
Performs cable compensation using known
capacitance source values.
Performs C-t measurements.
Places capacitance source values in a spreadsheet.
Read the cable compensation parameters and sends
them to the 590. This module is for internal use by the
SIMCVsweep82 and CTsweep82 modules. It is not
normally used as a stand-alone module.
Performs quasistatic measurement sweep.
Saves entered capacitance source values in a file.
Performs simultaneous C-V sweep.
D-29