Section 7: Keithley Configuration Utility (KCon)

System Configuration properties

Each instrument that is included in the system configuration is displayed in the Configuration
Navigator. When you select a Configuration Navigator node, the corresponding attributes or
properties are displayed in the Work Area. The following topics describe the properties for each
supported instrument.

KI 4200A SCS Properties

When you select KI 4200A SCS in the Configuration Navigator, the Work Area displays the system
configuration, as shown in the figure below.
System Properties area
The System Properties area of the Properties displays the 4200A-SCS serial number and other
software and platform information. You can change the
autorange method
Powerline frequency
Set Powerline Frequency to 60 Hz or 50 Hz. Make sure you set the powerline frequency correctly. If
the setting is wrong, the 4200A-SCS cannot properly reject powerline measurement noise.
7-12
Figure 413: KI System Configuration information
(on page 7-13), and
SMU standby range
Model 4200A-SCS Parameter Analyzer Reference Manual
Powerline frequency
(on page 7-13).