Loadcablecorrectionconstants - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual

LoadCableCorrectionConstants

LoadCableCorrectionConstants reads the cable compensation parameters for the range and frequency
specified from the cable compensation file and sends these parameters to the 590.
Usage
status = LoadCableCorrectionConstants(char *CabCompFile, char *instr_id,
*frequency, int range);
status
CabCompFile
instr_id
frequency
range
Details
If the file specified by CapCompFile does not exist, it is created. The path that you specify must
exist. When entering the path information, be sure to use two \ characters to separate each directory
level. For example, if your cable compensation file is in file C:\calfiles\590cal.dat, you would
enter C:\\calfiles\\590cal.dat.
range values
Range
1
2
3
4
The return values from status can be:
0: OK.
-10000 (INVAL_INST_ID): An invalid instrument ID was specified. This generally means that
there is no instrument with the specified ID in your configuration.
-10020 (COMP_FILE_ACCESS_ERR): There was an error accessing the cable compensation
file.
-10021 (COMP_FILE_NOT_EXIST): The specified compensation file does not exist.
-10022 (KI590_NOT_IN_KCON): There is no CMTR defined in your system configuration.
Also see
SaveCableCompCaps590 user module
4200A-901-01 Rev. C / February 2017
Returned values are placed in the Analyze sheet; see Details
The complete name and path for the cable compensation file; see Details
The CMTR instrument ID; this can be CMTR1 through CMTR4, depending on the
configuration of your system
The frequency of the correction constant: 0 = 1 MHz; 1 = 100 kHz
The range of the correct constant; see Details
100 kHz values
2 pF/2 μs
20 pF/20 μs
200 pF/200 μs
2 nF/2 ms
(on page B-34)
Appendix B: Using a Model 590 C-V Analyzer
1 MHz values
20 pF/200 μs
20 pF/200 μs
200 pF/2 ms
2 nF/20 ms
B-33