Keithley 4200A-SCS Reference Manual page 771

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 9: Keithley External Control Interface (KXCI)
PG
This command sets the trigger mode and initiates the start of pulse output or arms the pulse card.
Usage
PG A, B, C
A
B
C
Corresponding LPT library functions
pulse_trig
(on page 13-152)
pulse_burst_count
Details
The trigger mode setting affects both channels of a pulse card. For example, setting channel 1 of
pulse card 1 to Continuous also sets channel 2 to Continuous.
Example
PG 1, 2, 50
This command string sets pulse card 1 for the Trigger Burst trigger mode and a burst count of 50.
Also see
None
PH
This command stops all pulse output from the selected pulse card.
Usage
PH A
A
Corresponding LPT library functions
pulse_halt
(on page 13-143)
Example
PH 3
This command string stops pulse output from pulse card 2.
Also see
None
9-60
Pulse card channel: 1 to 8; the largest value is the number of channels in the
system
Trigger mode:
Burst mode: 0
Continuous: 1 (default)
Trigger burst: 2
Burst or trigger burst only: Pulse count in number of pulses: 1 to 2
continuous (default 1)
(on page 13-138)
Pulse card channel: 1 to 8; the largest value is the number of channels in the
system
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017
32
-1; set to 1 for

Advertisement

Table of Contents
loading