Model 4200A-SCS Parameter Analyzer Reference Manual

Fan operation

If the 4200A-SCS fan stops or is running too slowly, a warning message is displayed on the
instrument. If the fan does not return to normal within five minutes of the initial warning, the 4200A-
SCS is shut down to prevent system damage from overheating.
If the system was powered down because of a fan problem and is then powered up, another warning
message is displayed to indicate the reason for the shutdown.
If you see this message repeatedly, contact Keithley Instruments. See
(http://www.tek.com/service) for contact information.
For detail on fan operation, you can open the 4200A-SCS System Monitor. This is available in the
Windows system tray.
4200A-901-01 Rev. C / February 2017
Figure 492: 4200A-SCS System Monitor
Section 12: Maintenance
Service