Section 13: LPT library function reference

pulse_rise

This command sets the rise transition time for the pulse card pulse output.
Usage
int pulse_fall(int instr_id, long chan, double riset);
instr_id
chan
riset
Pulse modes
Standard
Details
Rise and fall transition time can be set independently for each pulse card channel. There is a
minimum slew rate for both the rise and fall transitions. For the fast speed range, the minimum is
362 µV/µs, or 1 V/2.7 ms. For the high-voltage range, the minimum slew rate is 1.8 mV/µs, or
1 V/500 µs. The pulse_range command is used to set pulse speed.
As shown below, the pulse rise time occurs between the 0 percent and 100 percent amplitude points
on the rising edge of the pulse, where the amplitude is the difference between the V High and V Low
pulse values.
The pulse rise time setting takes effect immediately during continuous pulse output. Otherwise, the
rise time setting takes effect when the next trigger is initiated. The pulse_trig command is used to
trigger continuous or burst output.
For slow speed, note that the minimum transition time for pulse source only (no measurement) on the
40 V range is 50 ns for the 4225-PMU and 4220-PGU.
Use the pulse_source_timing command to set the pulse fall time for the 4220-PGU and
4225-PMU.
Example
pulse_rise(VPU1, 1, 50e-9)
For fast speed, the sets the pulse rise time for channel 1 of the pulse card to 50 ns.
13-150
The instrument identification code of the pulse card, such as VPU1 or VPU2
Channel number of the pulse card: 1 or 2
Pulse rise time in seconds (floating point number):
Fast speed: 10e-9 to 33e-3
Slow speed, 4220-PGU and 4225-PMU: 50e-9 to 33e-3
Default: 100e-9