Keithley 4200A-SCS Reference Manual page 725

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 9: Keithley External Control Interface (KXCI)
CH
This command defines a SMU channel.
Usage
CHA, 'BBBBBB', 'CCCCCC', D, E
A
BBBBBB
CCCCCC
D
E
Details
For every used channel that is configured as a SMU, you must specify names for voltage and current,
select the source mode (voltage, current, or common), and select the source function (VAR1, VAR2,
constant, or VAR1').
When the source mode (D) is set to common (3), the source function (E) must be set to constant (3).
The VAR1 source function performs a linear or logarithmic sweep. The VAR1 function performs a
sweep that is synchronized to the steps of VAR2. The VAR1 sweep is performed whenever VAR2
goes to a new step value. The constant source function outputs a fixed (constant) source value.
The VAR1' source function is similar to the VAR1 function, except that each sweep step is scaled by
the Ratio value (RT) and an Offset (FS) as follows:
VAR1' sweep step = (VAR1 sweep step × RT) + FS
For example, assume VAR1 is set to sweep from +1 V to +3 V using 1 V steps. If Ratio (RT) is set to
2, and Offset (FS) is set to 1, each step of VAR1' is calculated as follows:
VAR1' step 1 = (1 V × 2) + 1 = 3 V
VAR1' step 2 = (2 V × 2) + 1 = 5 V
VAR1' step 3 = (3 V × 2) + 1 = 7 V
To disable a channel, send:
CHx
Where x = 1 to 9
Example 1
CH3, 'V1', 'I1', 1, 3
This command string sets up the SMU assigned to channel 3 to source a fixed voltage (1 V). The specified
names for voltage and current are V1 and I1, respectively.
9-14
1, 2, ... or n SMU channel number; the largest value of n is the number of channels in
the system (9 maximum); if no parameters are specified after the channel number,
the channel is turned off
Voltage name
Current name
Source mode or common:
1: Voltage source mode
2: Current source mode
3: Common (output high connected to common)
Source function:
1: VAR1 sweep source function
2: VAR2 step source function
3: Constant (fixed) source function
4: VAR1' source function
Model 4200A-SCS Parameter Analyzer Reference Manual