Keithley 4200A-SCS Reference Manual page 1226

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix F: Set up a probe station
Figure 627: Chuck movement
At the conclusion of a test, the site coordinates are recorded in the sheet settings. These coordinates
are only valid if a project uses the remote prober control (real prober). The coordinate system is
based on the xstart_position and ystart_position parameters of the prober-init action.
The site coordinates change only after a site movement is performed; the coordinates change when
the bottom of the project loop is reached. At the top of each iteration, the site coordinates remain the
same until the site movement is done. Refer to the following figure for an example of a table that
contains site coordinates (see column B, row 9).
4200A-901-01 Rev. C / February 2017
F-9