Section 6: Clarius
Set up Linear Segment Stress/Measure counts
To set counts for linear segment stress/measure mode:
1. Select Linear.
2. In First Stress Count, set the number of times that devices are stressed during the first segment.
3. In Total Stress Count, enter the total number of times the Segment Arb waveform will stress the
device. This is the total number of stress pulses that will be generated during all stress cycles.
4. In Number of Stresses, set the total number of stress cycles. This value must be less than the
value set for Total Stress Count.
5. If needed, enter the Segment Stress/Measure Delay in seconds. This is the delay after each
stress cycle. It allows the device to reach equilibrium before the next measurement.
Clarius uses the values you entered to calculate the segment stress counts for the segments. The
counts are displayed in the Stress Counts box.
6-172
Figure 332: Linear Segment Stress/Measure Mode
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017