Section 13: LPT library function reference
The 4225-PMU or 4220-PGU returns to the following states:
The pulse mode is maintained. For example, if the pulse card is in Segment Arb mode, it will still
be in Segment Arb mode after the devint process is complete.
5 V and 10 mA ranges
If in pulse mode:
Period of 1.0 µs
Transition Times (Rise and Fall) of 100 ns
Width of 500 ns
Voltage high and low of 0 V
Load of 50 Ω
If in segmented arb mode, Start Voltage is 0 V
If in arbitrary waveform mode, Table Length is 100
Also see
clrcon
(on page 13-199)
clrscn
(on page 13-9)
clrtrg
(on page 13-11)
dev_abort
(on page 13-92)
devclr
(on page 13-67)
kibdefint
(on page 13-24)

dsweepf

This command performs a dual frequency sweep.
Usage
int dsweepf(int instr_id, double startf, double stopf, long *NumPts, double
delaytime);
instr_id
startf
stopf
NumPts
delaytime
Details
Use the sweepf command to perform a single frequency sweep.
This command is used to perform a dual frequency sweep (see Example). The 4210-CVU provides
test frequencies from 1 kHz to 10 MHz in the following steps:
1 kHz through 10 kHz in 1 kHz steps
10 kHz to 100 kHz in 10 kHz steps
100 kHz to 1 MHz in 100 kHz steps
1 MHz to 10 MHz in 1 MHz steps
13-172
The instrument identification code of the 4210-CVU: CVU1
Initial frequency for the sweep
Final frequency for the first sweep
Variable to receive the number of points sourced during the sweep
Delay before each measurement (0 to 999 s)
Model 4200A-SCS Parameter Analyzer Reference Manual