Custom file arb waveforms (full arb)

The following figure summarizes the basic processes to create a custom full arb waveform file, to load
the file into a pulse card, and to output the pulse waveforms.
To create custom file arb waveforms (full arb):
1. Select and configure waveforms.
After you select an available waveform type, configure its settings, and name it, the waveform
is placed in the Scratch Pad.
Waveforms remain in the Scratch Pad until they are deleted by the user.
Refer to
for custom file arb. Refer to
page 10-10) for details.
2. Copy the waveforms into the Sequencer for Channel 1, Channel 2, or both:
The order that two or more waveforms appear in a channel sequencer is the order that the
waveforms are output by that channel.
Refer to
details.
3. Save the waveforms in the Sequencer as a .kaf file. See
waveforms into Sequencer
4. Load the .kaf waveform file into a pulse generator (using the appropriate pulse card tab): Refer
to
Custom Arb file operation: Load waveform and turn on output
5. Turn on the output for enabled channels. See
on output
4200A-901-01 Rev. C / February 2017
Figure 474: Basic process to create and output custom file Arb waveforms
Waveform types
(on page 10-13) for information about the waveform types available
Custom Arb file operation: Copy waveforms into Sequencer
(on page 10-11).
(on page 10-12).
Custom Arb file operation: Select and configure waveforms
Custom Arb file operation: Load waveform and turn
Section 10: KPulse (for Keithley Pulse Cards)
(on page 10-11) for
Custom Arb file operation: Copy
(on page 10-12) for details.
(on
10-9

Advertisement

Table of Contents
loading

Table of Contents