:CVU:BIAS:DCV:SAMPLE
This command configures the CVU to bias a DC voltage and sample n Z-measurements for the CVU card.
Usage
:CVU:BIAS:DCV:SAMPLE biasv, samples
biasv
samples
Details
Configures the CVU to bias a DC voltage and sample n Z-measurements for the CVU card. The other
parameters are set by their respective commands.
Also see
None
:CVU:SWEEP:FREQ
This command configures the CVU to sweep frequency and sample Z-measurements for the selected CVU card.
Usage
:CVU:SWEEP:FREQ fstart, fstop
:CVU:SWEEP:FREQ fstart, fstop, order
fstart
fstop
order
Details
Only the start frequency, stop frequency, and sweep order are set here. The other parameters are set
by their respective commands.
Values are coerced to one of the 37 discrete frequencies for the Model 4210-CVU.
Also see
None
4200A-901-01 Rev. C / February 2017
Voltage to source while sampling Z-measurements; voltages range from -30 V to
+30 V
The number of Z-measurements the CVU will make for the test operation; valid
values range from 1 to 4096
The frequency that is used for capturing the initial sample in the sweep
The frequency that is used for capturing the final sample in the sweep
Determines whether the test will bias DC voltage and sweep frequency or step
voltage and sweep frequency:
1: The voltage bias the voltage set by the :CVU:DCV command; the default
2: The voltage will step based on the vstart, vstop, vstep parameters set
by the :CVU:SWEEP:DCV command.