Section 9: Keithley External Control Interface (KXCI)
EC
This command sets the condition to exit the test if compliance is reached.
Usage
EC A
A
Example
EC 1
This command enables exit on compliance.
Also see
None
EM
This command switches between 4145 Emulation and 4200A command sets.
Usage
EM A,B
A
B
Example
EM 0,1
This selects the 4145 Emulation command set permanently.
Also see
None
9-48
Action on compliance:
Off (do not exit if compliance is reached): 0
On (exit if compliance is reached): 1
Mode:
4145 Emulation: 0
4200A: 1
Which sessions:
This session only: 0
This and all subsequent sessions (writes to KCon file): 1