Keithley 4200A-SCS Reference Manual page 794

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
GN
This command queries input or output parameter values, or both, by name for the last user module run in KXCI.
Usage
GN ParameterName
GN ParameterName NumValues
ParameterName
NumValues
Details
NumValues is only used for an output parameter that is an array. If NumValues is not used, one
value is returned.
Arrays are returned as a list of values separated by semicolons.
The value returned for an input parameter is the given value. The values returned for an output
parameter is the outcome of the test (for example, measured readings).
You can also query a parameter by specifying the corresponding number for the parameter in the
KULT module (see
Example
GN Vforce 11
This command queries all 11 test voltages (Vforce parameter) for the function that was run in the
page 9-82) example.
The following array of test voltages are returned and displayed in the KXCI console:
0; 0.5; 1.0; 1.5; 2.0; 2.5; 3.0; 3.5; 4.0; 4.5; 5.0
Also see
None
4200A-901-01 Rev. C / February 2017
The name of the parameter in the KULT module
The number of values in an output array to be returned; see Details
GP: get parameter (by number)
Section 9: Keithley External Control Interface (KXCI)
(on page 9-84)).
EX
(on
9-83

Advertisement