Section 13: LPT library function reference

mpulse

This command uses a source-measure unit (SMU) to force a voltage pulse and measure both the voltage and
current for exact device loading.
Usage
int mpulse(long instr_id, double pulse_amplitude, double pulse_duration, double
*v_meas, double *i_meas);
instr_id
pulse_amplitude
pulse_duration
v_meas
i_meas
Details
Voltage and current are measured just before the pulse terminates. Pulsing is useful for devices that
exhibit self-heating, which could damage the device or shift operating characteristics. Examples are
high-power GaAs transistors or BJTs, but even some silicon devices exhibit self-heating.
Example
double vdsat, idsat, vds;
.
.
mpulse(SMU1, vds, l.0E-3, &vdsat, &idsat);
/* Pulse output of SMU1. */
This example measures the drain current of a metal-oxide semiconductor field-effect transistor (MOSFET)
when drain-source voltage (V
drain. The pulse duration is 1 ms. Voltage across the MOS transistor, V
measured.
Also see
None
13-80
The instrument identification code of the instrument under control
The pulse height in volts
The pulse width in seconds; the measurements are made at the end of the pulse
before the mpulse command is shut down
The variable used to receive the voltage on the output of the instrument at the time
the pulse terminates
The variable used to receive the current drawn from the instrument; this
measurement is made simultaneously with the voltage, so the combined values are
an exact representation of the device load at pulse termination
) equals gate-source voltage (V
DS
Model 4200A-SCS Parameter Analyzer Reference Manual
). A voltage pulse, V
GS
, and drain current, I
DSAT
4200A-901-01 Rev. C / February 2017
, is applied to the
DS
, are
DSAT