Keithley 4200A-SCS Reference Manual page 1121

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Appendix B: Using a Model 590 C-V Analyzer
Csize
V
Vsize
G_or_R
G_or_Rsize
T
Tsize
Details
This user module performs a capacitance versus voltage staircase sweep. The figure below shows
the default parameters for the 590-cvsweep UTM, which uses the CvSweep590 user module.
B-28
Set to a value equal to or greater than the number of voltage steps in the sweep or
equal to ((LastBias − FirstBias) / StepV) + 1
Output: The array of bias voltages used
Set to a value equal to or greater than the number of voltage steps in the sweep or
equal to ((LastBias − FirstBias) / StepV) + 1
Output:
When the parallel measurement model (0) is selected, G_or_R is the
measured conductance
When the series measurement model (1) is selected, this is the measured
resistance
When this function is called from a Clarius UTM, this parameter is fixed at 1350
The array of time stamps for each measurement step
When this function is called from a Clarius UTM, this parameter is fixed at 1350
Figure 566: CvSweep590 user module (590-cvsweep UTM)
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017