Keithley 4200A-SCS Reference Manual page 757

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 9: Keithley External Control Interface (KXCI)
BC
This command clears all readings from the buffer.
Usage
BC
Details
The GPIB data buffer can hold up to 4096 readings.
This command string clears all readings from the buffer. It also clears bit B0 (Data Ready) of the
status byte.
Also see
None
RS
This command sets the measurement resolution for all channels.
Usage
RS A
A
Example
RS 7
This command string provides full SMU resolution when the 4200A command set is selected.
Also see
None
RI
This command instructs a SMU to go a specified range immediately without waiting until the initiation of a test.
Usage
RI channel,range,compliance
channel
range
compliance
Example
RI 1, 1E-3, 1E-3
This command string instructs SMU1 to go to the 1 mA range and set compliance to 1 mA.
Also see
None
9-46
Resolution, in number of digits:
4200A command set: 3 to 7
4145 Emulation command set: 3 to 5
SMU number (1 to 9)
Range, permitted values:
4200-SMU without a preamplifier: 100e-9 A to 100e-3 A
4200-SMU with a preamplifier: 1e-12 A to 100e-3 A
Compliance, permitted values: 10% to 100% of range
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017