Model 4200A-SCS Parameter Analyzer Reference Manual

PguInit8110 user module

This user module initializes the pulse generator to a default setup.
Usage
status = PguInit8110(char *instr_id);
status
instr_id
Details
The user module used by the pgu1-init UTM.
The PguInit8110 user module initializes the Keysight 8110A pulse generator as follows:
Disables the output of the specified channel.
Resets (*RST) to ensure that all errors are cleared.
Sets the output polarity to NORMAL.
Sets the trigger count to 1.
Sets the trigger source to MANUAL.
Enables SINGLE PULSE mode.
Allows the rise/fall to be independently programmable.
Sets the pulse height to 0.2 V and base to 0 V.
Sets the rise/fall to 100e-9 s.
Sets the width to 300e-9 s.
Disables error checking.
Also see
None
4200A-901-01 Rev. C / February 2017
Returned values are placed in the Analyze sheet:
0: OK
-10000 (INVAL_INST_ID): The specified instrument ID does not exist
-10040 (HP8110_NOT_IN_KCON): No PGU was found in the system
configuration
-10041 (HP8110_NOT_INITED): The PGU was never initialized
-10042 (HP8110_PULSE_ERROR): There was an error during pulsing
-10090 (GPIB_ERROR_OCCURRED): A GPIB communications error
occurred
-10091 (GPIB_TIMEOUT): A time-out occurred during communications
-10100 (INVAL_PARAM): An invalid input parameter is specified
The PGU (pulse generator) instrument ID: PGUX, where X is a number from 1
through 8 (configuration dependent); the PGU instrument ID effectively corresponds
to a single pulse generator channel
Appendix E: Using a Keysight 8110A/8111A Pulse Generator
E-5