Measurement Category II (as referenced in IEC 60664) connections require protection for high transient overvoltages often associated with local AC mains connections. Certain Keithley measuring instruments may be connected to mains. These instruments will be The following safety precautions should be observed before using this product and any marked as category II or higher.
Page 3
If you are using a test fi xture, keep the lid closed while power is applied to the device under selected parts should be purchased only through Keithley Instruments to maintain test. Safe operation requires the use of a lid interlock.
Power and environmental specifi cations For indoor use only. Carefully consider and confi gure the appropriate Power Supply 100 VAC to 240 VAC, 50 Hz to 60 Hz output-off state, source levels, and compliance Maximum VA 1000 VA levels before connecting the instrument to a device that can deliver energy.
Page 5
Introduction USB fl ash drive contents The easy-to-use Model 4200A-SCS Parameter Analyzer The USB fl ash drive that is included with your instrument contains performs lab grade DC and pulse device characterization, installation fi les for Clarius+. These can only be installed on the real-time plotting, and analysis with high precision and 4200A-SCS Parameter Analyzer.
(60 pounds). Do not lift the 4200A-SCS Remove the documentation, standard accessories, USB alone and do not lift the 4200A-SCS using the fl ash drive, and optional accessories (such as rack-mount front bezel. Lifting the instrument by the front hardware).
Page 7
You should have received the 4200A-SCS Parameter Analyzer with: • Keyboard with integrated mouse • Clarius+ software USB fl ash drive • Power line cord • 236-ILC-3 safety interlock cable • 4200A-SCS Quick Start Guide (not shown; this document) Refer to the packing list for additional items that might have shipped with your instrument.
42.4 V peak, or 60 VDC for equipment rated for dry disconnect power sources when a test fi xture cover is opened. locations. Keithley Instruments products are only rated for • Where possible, use automatic handlers so operators are not dry locations.
Install the instrument If you need voltages greater than ±40 V, you must also connect The 4200A-SCS can be used on a bench or in a rack. Please the exterior of the test fi xture enclosure to protective earth see the instructions that came with your rack-mount kit if you are (safety ground).
Page 10
High-voltage test fi xtures require extra precaution to ensure there are no shock hazards. For correct operation with the 4200A-SCS, the test fi xture should have a normally open switch that is used for the interlock. An open interlock condition occurs when the switch is open.
Connect one end of the supplied interlock cable (part If you are testing discrete devices, you need a test fi xture that is number 236-ILC-3) to the rear panel of the 4200A-SCS. equipped with 3-lug triaxial connectors. The 8101-PIV test fi xture,...
Page 12
Power on the 4200A-SCS The 4200A-SCS operates from a line voltage of 100 V to 240 V at a frequency of 50 Hz or 60 Hz. Make sure the operating voltage in your area is compatible. The power cord supplied with the Model...
Page 13
“Yes” to an on-screen license agreement. Answering “No” AC line power receptacle. makes your system nonfunctional until you reinstall the software. Turn on the 4200A-SCS by pushing the power switch to the lit position. The instrument starts up. FAQs Safety...
Page 14
Close Clarius. Run KCon. The default powerline frequency is set to 60 Hz. If the setting is wrong, the 4200A-SCS cannot properly reject powerline Select KI 4200A SCS in the Confi guration Navigator. The left pane is the Confi guration Navigator.
Page 15
FAQs Safety Introduction Connect Next steps...
Page 16
Do not spray cleaning fl uids onto the communication problems between instruments and the display. Additionally, it is strongly recommended that you use only 4200A-SCS. fi ngers on the touchscreen. Do not use sharp objects, such as a pen or pencil, to touch the touchscreen.
Next steps • Data sheets: Technical data regarding the 4200A-SCS and See the 4200A-SCS Learning Center , which is preinstalled related accessories. on your system. To access the Learning Center, click Help in the Clarius menu, press F1 while using Clarius, or select the desktop See: www.tek.com/keithley...