Keithley 4200A-SCS Reference Manual page 370

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
Waveform Capture (Discrete Pulses) measurement example
For the example shown in the next figure, the samples of three pulses are captured. The 51 samples
(17 samples x 3 pulses) are placed in the Analyze sheet.
Figure 266: Waveform capture - Discrete pulses
Number of Pulses
The number of pulses for the PMU channel to output and measure for each step in the sweep (1 to
10,000). If you enabled autorange, load-line effect compensation (LLEC), or thresholds, the number
of pulses is output multiple times for each step in a sweep.
Timing Sweep
You can only step the timing parameters if the Operation Mode is set to pulse sweep or pulse train
(not pulse step) and if Test Mode is set to Pulse IV.
You can only sweep the timing parameters if the Operation Mode is set to pulse step or pulse train
(not pulse sweep).
The Timing Sweep Step option is disabled when Waveform Capture is selected.
The next figure shows parameter stepping and sweeping examples. In these examples, the Number
of Pulses parameter is set to 3, so there are three pulses per period.
Figure 267: Parameter stepping sweeping examples
4200A-901-01 Rev. C / February 2017
6-101

Advertisement

Table of Contents
loading

Table of Contents