Code Examples - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 9: Keithley External Control Interface (KXCI)
:CVU:CONFIG:ACVHI
This command allows you to define the source terminal (AC only) for the CVU test.
Usage
:CVU:CONFIG:ACVHI source
source
Also see
None
:CVU:CONFIG:DCVHI
This command allows you to define the source terminal (DC only) for the CVU test.
Usage
:CVU:CONFIG:DCVHI source
source
Also see
None

Code examples

Example 1
The following code segment sets CVU1 to perform a system mode sweep of DC voltage from 5 V to
10 V in 1 V steps. After the test completes, the Z, DCV, F, timestamps, and status values are queried.
The operation mode is then changed to perform a list sweep of DC voltage (2 V, 4 V, 3 V, 5 V, 7 V)
and the test is run again with all other test conditions as previous.
// Soft Reset card
send(addr, ":CVU:RESET", &status);
// Set CVU to System Mode
send(addr, ":CVU:MODE 1", &status);
// Set measurement model to Z, theta
send(addr, ":CVU:MODEL 0", &status);
// Set speed to Normal
send(addr, ":CVU:SPEED 1", &status);
// Set AC drive level to 45 mV at 7MHz
send(addr, ":CVU:ACV 0.045", &status);
send(addr, ":CVU:FREQ 7E+6", &status);
// Set DC bias level to 10 V
send(addr, ":CVU:DCV 10", &status);
9-78
The source terminal to be used:
HCUR/HPOT (default): 1
LCUR/LPOT: 2
The source terminal to be used:
HCUR/HPOT (default): 1
LCUR/LPOT: 2
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017

Advertisement