Gpib Connections; Using Kcon To Add A Keysight Lcr Meter To The System - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
2-wire remote sensing
For 2-wire local sense connections, coaxial tees are required to adapt dual BNC cables to single BNC
cables, as shown in the following figure.
Figure 574: 2-wire local sense connections to equipment using triaxial connectors

GPIB connections

The 4200A-SCS controls the Model 4284A or 4980A through the General Purpose Interface Bus
(GPIB). Use the Keithley 7007-1 or 7007-2 GPIB cable to connect the GPIB port of the Model 4284A
or 4980A to the GPIB port of the 4200A-SCS.

Using KCon to add a Keysight LCR Meter to the system

To use the 4200A-SCS to control an external instrument, you must add the instrument to the 4200A-
SCS system configuration. You use Keithley Configuration Utility (KCon) to add the Keysight Model
4284A or 4980A to the test system.
Refer to
For additional details on KCon, refer to
4200A-901-01 Rev. C / February 2017
Using KCon to add equipment to the 4200A-SCS
Appendix C: Using a Keysight 4284/4980A LCR Meter
Figure 575: Equivalent circuit
(on page 7-7) for instruction.
Keithley Configuration Utility
(on page 7-1).
C-5