Model 4200A-SCS Parameter Analyzer Reference Manual
To open and execute the QTsweep UTM:
1. Choose Select.
2. In the Test Library, select Custom Test, Choose a test from the pre-programmed library
(UTM).
3. Drag the test to the project tree.
4. Select Configure.
5. In the Test Settings pane, under User Libraries, select ki82ulib.
6. Under User Modules, select QTsweep82.
7. Modify the test parameters as needed. Refer to
parameter definitions.
If you use the parameters shown in the figure below, the Model 82 makes 20 quasistatic
capacitance measurements using 20 mV pulses (V_Step) ranging from 0.07 seconds to
1 second (Delay_Max).
8. Click Run to execute the test.
4200A-901-01 Rev. C / February 2017
QTsweep82 user module