Model 4200A-SCS Parameter Analyzer Reference Manual

Data format for user mode readings

For user mode operation, use the TI or TV command string to trigger and make a reading. After
sending the TI or TV command string and addressing the 4200A-SCS to talk, the output data string is
sent to the computer in the following format:
XYZ ±N.NNNN E±NN
Data status
The hierarchy for data status is L, V, X, C, T, N, where X is:
N: Normal
L: Interval too short
V: Overflow reading (A/D converter saturated)
X: Oscillation
C: This channel in compliance
T: Other channel in compliance
Measure channel
Voltage measure mode specified (Z = V; see below)
Y =
Current measure mode specified (Z = I; see below)
Y =
4200A-901-01 Rev. C / February 2017
A
SMU1
B
SMU2
C
SMU3
D
SMU4
VM1
E
F
VM2
G
SMU5
SMU6
H
I
SMU7
SMU8
J
K
VM3
L
VM4
VM5
M
N
VM6
VM7
O
P
VM8
SMU1
A
B
SMU2
C
SMU3
D
SMU4
E
SMU5
SMU6
F
G
SMU7
H
SMU8
Section 9: Keithley External Control Interface (KXCI)