Appendix E: Using a Keysight 8110A/8111A Pulse Generator
Also see
PguTrigger8110

PguTrigger8110 user module

This user module specifies number of pulses to output and triggers the start of the pulse output
process.
Usage
status = PguTrigger8110(char *InstIdStr, double Count);
status
InstIDStr
Count
Details
The PguTrigger8110 function will trigger the pulse (or pulses) defined using the PguSetup8110
function.
Also see
PguSetup8110 user module
E-8
(on page E-8)
Returned values are placed in the Analyze sheet:
0: OK
-10000 (INVAL_INST_ID): The specified instrument ID does not exist
-10040 (HP8110_NOT_IN_KCON): No PGU was found in the system
configuration
-10041 (HP8110_NOT_INITED): The PGU was never initialized
-10042 (HP8110_PULSE_ERROR): There was an error during pulsing
-10090 (GPIB_ERROR_OCCURRED): A GPIB communications error
occurred
-10091 (GPIB_TIMEOUT): A timeout occurred during communications
-10100 (INVAL_PARAM): An invalid input parameter is specified
The PGU (pulse generator) instrument ID: PGUX, where X is a number from 1
through 8 (configuration dependent); the PGU instrument ID effectively corresponds
to a single pulse generator channel
The number of pulses to output; if Count is > 1, a burst of pulses with a period as
defined in the PguSetup8110 function is output; if Count is 1, a single pulse is
output
(on page E-6)
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017