Pulse_Trig_Output - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
Also see
pulse_burst_count
pulse_delay
pulse_halt
pulse_output
pulse_trig_source

pulse_trig_output

This command sets the output trigger on or off.
Usage
int pulse_trig_output(int instr_id, long state);
instr_id
state
Pulse modes
Standard, Full Arb, Segment Arb
Details
This command turns the TTL-level trigger output pulse on or off. The pulse is used to synchronize
pulse output with the operations of an external instrument. When connected to a scope, each output
pulse of triggers a scope waveform measurement.
When output trigger is enabled, an output pulse will initiate a TTL-level, 50% duty cycle output trigger
pulse. The trigger pulses are available at the TRIGGER OUT connector of the pulse generator card.
The figure below shows the behavior of output triggers (T
the Burst mode, output triggers continue even though pulse output has stopped. For the trigger burst
mode, output triggers stop when the pulse output stops.
4200A-901-01 Rev. C / February 2017
(on page 13-138)
(on page 13-141)
(on page 13-143)
(on page 13-146)
(on page 13-155)
The instrument identification code of the pulse card, such as VPU1 or VPU2
Output trigger state:
Off: 0 (default for Segment Arb and full arb)
On: 1 (default for standard pulse)
Figure 511: Pulse generator card output trigger
Section 13: LPT library function reference
) for the three trigger modes. Notice that for
O
13-153