Forcev - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
Also see
asweepv
dsweepf
(on page 13-172)
setfreq
(on page 13-181)
setlevel
(on page 13-182)
sweepf
(on page 13-192)
sweepv
(on page 13-193)

forcev

This command sets the DC bias voltage level.
Usage
int forcev(int instr_id, double voltage);
instr_id
voltage
Details
This command sets a DC bias level for a single impedance measurement and a frequency sweep.
Use the setfreq and setlevel commands to set the AC drive frequency and AC voltage for the
sweep.
The DC source operates independently of the AC source. Changes to the level and state of the DC
source take effect immediately; the AC frequency and source value are only used during measz
operations.
Example
Programming example #1
rdelay command provides a settling time before the measurement.
Also see
measz
(on page 13-178)
setfreq
(on page 13-181)
setlevel
(on page 13-182)
4200A-901-01 Rev. C / February 2017
(on page 13-167)
The instrument identification code of the 4210-CVU: CVU1
DC bias voltage level (−30 V to 30 V)
(on page 13-194) makes a single impedance measurement. Note that the
Section 13: LPT library function reference
13-175