Keithley 4200A-SCS Reference Manual page 1003

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 13: LPT library function reference
The pulse voltage high setting takes effect immediately during continuous pulse output. Otherwise,
the voltage high setting takes effect when the next trigger is initiated. The pulse_trig command is
used to trigger continuous or burst output.
The pulse_vlow, pulse_vhigh, and pulse_dc_output commands set the voltage value
output by the pulse channel when it is turned on (using pulse_output). If the output is already
enabled, these commands change the voltage level immediately, before the pulsing is started with a
pulse_trig command.
Example
pulse_vhigh(VPU1, 1, 2.5)
Sets the pulse voltage high value for channel 1 of the pulse card to 2.5 V.
Also see
pulse_dc_output
pulse_output
(on page 13-146)
pulse_range
(on page 13-149)
pulse_trig
(on page 13-152)
pulse_vlow
(on page 13-159)
13-158
(on page 13-140)
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017

Advertisement

Table of Contents
loading

Table of Contents