User Mode Commands (Us) - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
Fixed source ranging command
You can use the following command with any system mode page.
SR
This command sets a fixed source range on channel n.
Usage
SR A, B
A
B
Details
The default setting is autorange for backward compatibility. If you specify a range that is below the
bias or sweep parameters that follow, the range is adjusted to accommodate the sweep.
Example
SR 1, 2
This command string selects best fixed range on channel 1.
Also see
None

User mode commands (US)

You can use the user mode (US) command strings for the following operations:
SMU setup
VS1...VS9 setup
Triggering
To send these command strings to the 4200A-SCS, select the user mode by sending the command:
US
4200A-901-01 Rev. C / February 2017
The channel to be controlled; a value between 1 and the number of channels in the
system (9 maximum)
Range:
Auto: 0
Best fixed range (determined by maximum sweep parameters: 2
Fixed range: > 0 to 1.0