Section 10: KPulse (for Keithley Pulse Cards)

Standard pulse waveforms

Standard pulse waveforms are configured and controlled from the pulse card tabs in the KPulse
window. The following figure explains how to use KPulse for a standard pulse output.
Standard pulse waveform previewers: KPulse provides a preview of configured standard pulse
waveforms for each enabled channel. Each waveform previewer shows the high and low levels, and
timing for the waveform.
In the next figure, the configuration shown in the waveform previewer for Channel 1 uses the default
settings for KPulse (pulse high = 1 V and pulse low = 0 V). Channel 2 uses the same settings, but the
complement mode is enabled. Pulse high goes to pulse low level (0 V) and pulse low goes to pulse
high level (1 V).
The following instructions describe how to use KPulse for a standard pulse output. Repeat these
steps for any installed cards that you want to set to standard pulse.
10-4
Model 4200A-SCS Parameter Analyzer Reference Manual
Figure 471: Standard pulse operation settings