Section 13: LPT library function reference

cvu_custom_cable_comp

This command determines the delays needed to accommodate custom cable lengths.
Usage
cvu_custom_cable_comp(int instr_id);
instr_id
Details
The custom cable length measure gathers a specific set of timing coefficients to be applied during the
C-V testing for a custom length cable. They are used to compensate the calibrated measurements
made from the CVU.
Custom cable lengths are any lengths that are not 0 m, 1.5 m, or 3 m.
Once this command is run, these values are applied if you select a cable length of Custom in Clarius
Tools > CVU Connection Compensation.
Possible return values are:
0: OK
-907: LPOT/LCUR fail
-908: HPOT/HCUR fail
Also see
Connection compensation

devclr

This command sets all sources to a zero state.
Usage
int devclr(void);
Details
This command clears all sources sequentially in the reverse order from which they were originally
forced. Before clearing all Keithley supported instruments, GPIB-based instruments are cleared by
sending all strings defined with the kibdefclr command.
devclr is implicitly called by clrcon, devint, execut, and tstdsl.
For C-V testing, this command turns off the DC bias voltage.
Also see
clrcon
(on page 13-199)
devint
(on page 13-13)
execut
(on page 13-16)
kibdefclr
(on page 13-22)
tstdsl
(on page 13-50)
13-170
The instrument identification code of the CVU (CVU1)
(on page 4-10)
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017