Section 9: Keithley External Control Interface (KXCI)
DV and DI
These commands set up each SMU channel.
Usage
AAB, CC, ±DDD.DDDD, ±EEE.EEEE
AA
B
CC
DDD.DDDD
EEE.EEEE
9-40
Source mode:
•
Voltage source: DV
•
Current source: DI
SMU channel: 1 to 9
Voltage source range:
•
Autorange: 0
•
20 V range: 1
•
200 V range: 2
•
200 V range: 3
•
200 mV range: 4 (only with a preamplifier)
•
2 V range: 5 (only with a preamplifier)
Current source range:
•
Autorange: 0
•
1 nA range (only with a preamplifier): 1
•
10 nA range (only with a preamplifier): 2
•
100 nA range: 3
•
1 µA range: 4
•
10 µA range: 5
•
100 µA range: 6
•
1 mA range: 7
•
10 mA range: 8
•
100 mA range: 9
•
1 A range (only with a 4210-SMU): 10
•
1 pA range (only with a preamplifier): 11
•
10 pA range (only with a preamplifier): 12
•
100 pA range (only with a preamplifier): 13
Output value:
•
Volts: -210.00 to +210.00
•
Amps, 4200-SMU: -0.1050 to +0.1050
•
Amps, 4210-SMU: -1.0500 to +1.0500
Compliance value:
•
Volts: -210.00 to 210.00
•
Amps, 4200-SMU: -0.1050 to +0.1050
•
Amps, 4210-SMU: -1.0500 to +1.0500
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017