Model 4200A-SCS Parameter Analyzer Reference Manual
This figure shows an example of a Segment Arb sequence defined by the seg_arb_sequence
command. Spot mean discrete measurements are performed on segments two and four.
This table lists the seg_arb_sequence parameter arrays for the Segment Arb sequence shown in
the example.
Parameter arrays for the seg_arb_sequence example
Parameter
SegNum
StartV
StopV
Time
Trig
SSR
MeasType
MeasStart
MeasStop
Also see
Segment Arb waveform
seg_arb_waveform
4200A-901-01 Rev. C / February 2017
Figure 508: Segment Arb sequence example
Array Name
Value
Seg_Num
1
Start_Volt
0 V
Stop_Volt
1 V
Time_Interval
50e-9 s
Trigger_Level
1 (high)
Output_Relay
1 (closed)
Meas_Type
0 (none)
Meas_Start
0 s
Meas_Stop
0 s
(on page 5-56)
(on page 13-128)
2
3
4
1 V
1 V
1.5 V
1 V
1.5 V
1.5 V
100e-9 s
20e-9 s
150e-9 s
1 (high)
1 (high)
0 (low)
1 (closed)
1 (closed)
1 (closed)
1 (spot mean) 0 (none)
1 (spot mean)
25e-9 s
0 s
50e-9 s
75e-9 s
0 s
100e-9 s
Section 13: LPT library function reference
5
6
1.5 V
0 V
0 V
0 V
50e-9 s
500e-9 s
0 (low)
0 (low)
1 (closed)
1 (closed)
0 (none)
0 (none)
0 s
0 s
0 s
0 s
7
0 V
0 V
130e-9 s
0 (low)
0 (open)
0 (none)
0 s
0 s