Model 4200A-SCS Parameter Analyzer Reference Manual
3. Select Probe Station.
4. Select OK. The KCon navigation displays PRBR1 in the Configuration Navigator and the
properties.
5. For the Model, select the Signatone CM500 (WL250) Prober.
6. Make sure the Number of Pins / Positioners is correct. The number of pins defined here
determines the pins that are available to assign to a switch matrix card column.
7. Save the configuration.
8. Exit KCon.
Clarius project example
To set up a new prober project:
1. Start Clarius.
2. Choose Select.
3. Select the Projects tab.
4. Drag New Project into the project tree.
5. Choose Yes to create a new project.
6. Rename the project PRB_CM500.
7. Select the Wafer Plan tab.
8. Drag Site to the project tree.
9. Select the Actions tab.
10. In the Search box, enter prober.
11. Drag the Prober Initialization (prober-init) action to the project tree. Make sure it is under the
subsite.
4200A-901-01 Rev. C / February 2017
Figure 820: Use KCon to select a prober
Appendix K: Using a Signatone CM500 Prober