Data Output
Saving and Recalling Instrument State

Recall Procedure Using "Recall by File Name" Feature

You can use the recall feature with the
named freely and saved in the D:\State folder. This function lets you recall a file you have
named freely and saved by simple softkey operation, eliminating annoying operation using
the Open dialog box.
NOTE
Although there is no limit to the number of files saved in a folder, only up to 50 files are
displayed on the softkeys. If more than 50 files are saved in a folder, they are sorted in the
order of numbers 0 to 9 and alphabetic characters A to Z and the first 50 files are displayed
as softkeys.
Although there is no limit to the number of characters of a file name, only up to 12
characters are displayed on the softkey. If a file name exceeds 12 characters, the first 12
characters are displayed on the softkey and the remaining characters are omitted and
replaced with "...".
Note that different files may be displayed on softkeys with the same name or a saved file is
not displayed on any softkey because of the above limitations.
Step 1. Press
Step 2. Press
Step 3. Files that have been named and saved in the D:\State folder are displayed on softkeys.
Press the key for the file you want to recall.

Priority of recalling the configuration file at startup

If several instrument configuration files exist at the startup of the E5070B/E5071B, only
one file is recalled and set at a time in the following order of priority.
If these files do not exist, the normal preset (factory preset) is executed.
Priority
1
2
3
358
.
.
Recall by File Name
Recalled file
Configuration file for the auto-recall function in the A drive
Configuration file for the auto-recall function in the D drive
Configuration file for the user-preset function in the D drive
*1.Executed when the preset operation mode is User and the file (D:\UserPreset.sta) exists.
softkey for files you have
Recall by File Name
*1
Chapter 10