Data Output

Saving and Recalling Instrument State

Saving and Recalling Instrument State
Using the Save/Recall functions, it is possible to save instrument state from the
E5070A/E5071A to a storage medium (e.g., hard disk or floppy disk) for future use. Table
9-1 lists the available save contents.
Table 9-1
Save Contents
Save settings (State
Only)
Save settings and
calibration data
(State & Cal)
Save settings and
traces (State &
Trace)
Save settings,
calibration data,
and traces (All)

File Compatibility in Save/Recall Operations

Unlike saving/recalling data from the same unit of E5070A/E5071A, following
compatibility issues must be considered when transferring data between multiple
E5070A/E5071A units.
1. Exchanging files between different models
2. Exchanging files between units with different options
NOTE
Recalling an incompatible file will result in an error and the settings will be reset.
196
Save type
Saves the settings
later restore the operating environment of the E5070A/E5071A.
Saves the settings
*2
array
). By recalling the saved settings, you can later restore the operating
environment of the E5070A/E5071A. This option also allows the user to make error
corrections using the recalled calibration data.
Saves the settings
memory array
environment of the E5070A/E5071A. Saved traces are also recalled together with the
settings and displayed on the screen.
Saves the settings
saved settings, you can restore the operating environment of the E5070A/E5071A.
Saved calibration data and traces are also recalled together with the settings.
*1.See Appendix C, "List of Default Values," for settings to be saved.
*2.See "Data Processing" on page 456 for details about arrays.
It is not possible to exchange files between different models, e.g., a file saved by the
E5070 cannot be loaded into the E5071A.
"State Only" files:
A unit equipped with more ports can load a file saved from a unit with less ports.
However, the reverse is not possible.
A unit with a source attenuator can load a file saved from a unit with no source
attenuator. However, the reverse is not possible.
File types other than "State Only":
File exchange is not possible if the number of ports and the presence of a source
attenuator does not match between the units.
Save contents and purposes
*1
on the E5070A/E5071A. By recalling the saved settings, you can
*1
of E5070A/E5071A and calibration data (calibration coefficient
*1
E5070A/E5071A and traces (corrected data array
*2
). By recalling the saved settings, you can later restore the operating
*1
of E5070A/E5071A, calibration data, and traces. By recalling the
*2
and corrected
Chapter 9