Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
When Something Goes Wrong
"Correlation Invalid, Module Did Not Trigger" on page 38
"Correlation Invalid, Timer Overflowed" on page 38
"Measurement Error: Check Probe Parametrics or Grounding" on page 38
"Clock Lines Must Be Connected to the Card in Slot X" on page 37

Maximum of 32 Channels Per Label

The logic analyzer can only assign up to 32 channels for each label. If
you need more than 32 channels, assign them to two labels and use the
labels in conjunction.
See "Adding and Deleting Labels for Terms" on page 98 for how to use
more than one label with trigger terms.

Measurement Initialization Error

The logic analyzer module failed the internal calibration which it
performs when Run is selected. An internal calibration failure can
indicate either a software or a hardware problem.
Possible Causes
• Hardware failure
• Software failure
Run the Self-Test Utility (see page 40) on the logic analyzer and
contact your Agilent Technologies Sales Office for service or software
upgrades.

Slow or Missing Clock

The message "Slow or Missing Clock" only appears in state
measurements. However, if you have another instrument armed by the
state analyzer, a slow or missing clock on the state analyzer will prevent
the other instrument from triggering also.
Possible Causes
• Target system is not running properly
34