10 Maintenance
10.11 Test exposures/Test images
452
7. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if it has been configured).
The exposure is displayed in the exposure window.
For notes on the evaluation of the exposure, refer to technical
documentation for acceptance/constancy testing.
8. Remove the 3D test phantom again.
10.11.3.6

Test image (3D)

You can use the
and the operability of the flat panel detector.
1. Call up the menu
2. Select the
control exposures"
In the action area, the menu appears to create the
3. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
"Test image"
menu to check the 3D data path of the unit
"Control exposures"
"Test image"
element in the structure tree under
.
"Start acquisition"
button.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
[ → 428].
"3D
"Test image"
64 95 258 D3632
D3632.076.01.03.02 09.2016
.