Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
SR*
S008
S008.2
S008.3
64 95 258 D3632
D3632.076.01.03.02 09.2016
8.1.8
S008: Update service
SA**
Function
Update service
No
Overview of module software versions
No
Confirming the unit serial number
* SR=service routine, ** SA=security access
8.1.8.1
S008: Test step 2
Overview of module software versions
1. Call service routine S008.2 [ → 249].
2. Exit the service routine [ → 252].
8.1.8.2
S008: Test step 3
Confirming the unit serial number
IMPORTANT
If the backup copy of the old unit serial number does not match the new
one after replacing a module, the entry of the serial number is activated.
If an incorrect serial number is entered, the message "FFFF" appears
on the display. In this case, the service routine can be run again.
1. Call up service routine S008.3 [ → 249].
2. Confirm the serial number displayed by pressing the R key (A).
3. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
The software versions currently installed on the modules are
displayed on an info screen on the touchscreen display.
8 Service routines
267