8 Service routines
8.1 Perform service routines via the control panel
SR*
S005
S005.2
S005.4
S005.5
S005.6
S005.7
260
8.1.6

S005: General X-ray tube assembly service

SA**
Function
General X-ray tube assembly service
yes
Automatic adjustment of preheating
no
Fan test
no
Temperature sensor test, single tank
yes
Switch off cool-down interval of the single tank
no
Configuring continuous operation of the tube assembly fan
* SR=service routine, ** SA=security access
8.1.6.1

S005: Test step 2

Automatic adjustment of preheating
Unit is radiating X-rays.
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
IMPORTANT
The tube assembly must be at operating temperature before the
compensation. To this end, run service routine S001 to release radiation
once for 14.0 s at kV/mA level 60/8 [ → 253].
1. Call service routine S005.2 [ → 249].
2. Start automatic adjustment by releasing radiation.
Important: Keep pressing the release button until adjustment is
completed and the new offset value for preheating is displayed. If you
interrupt the adjustment procedure prematurely by letting go of the
release button, the message "EEEE" appears in selection field 3. This
message may have to be acknowledged by pressing the R key.
WARNING
An inactive progress indicator in selection field 1 and the
message "FFFF" in selection field 3 signal that the system is
ready for compensation.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016