2. Use the arrow keys (A) in the corresponding selection field to select
the required offset for the switching point (see table).
IMPORTANT
A negative offset shifts the respective switching point toward larger
skulls; a positive offset has the opposite effect.
3. Save the setting [ → 251].
4. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
8 Service routines