5 Messages
5.4 List of error messages
Error code
Description
E6 22 01
DCS sensor could not be initialized. ● Check correct positioning of the DCS
Error code
Description
E6 22 03
Error during communication to DCS
sensor.
Error code
Description
E6 22 10
No image data obtained from DCS
sensor or interruption of the data
flow.
Error code
Description
E6 22 15
Serial number of the DCS sensor
could not be read out.
Error code
Description
E6 22 16
File operation malfunction in DCS
sensor mode.
Error code
Description
E6 22 17
Tolerance exceeded when
monitoring the high-voltage of the
DCS sensor.
118
5.4.10

Location 22: DCS sensor

Actions required
E6 22 01
Actions required
E6 22 03
● Check correct positioning of the DCS
Actions required
E6 22 10
● Check correct positioning of the DCS
Actions required
E6 22 15
If the error does
errors ...
● Replace the DCS sensor.
Actions required
E6 22 16
If the error occurs repeatedly ...
● Replace the DCS sensor.
Actions required
E6 22 17
If the error occurs repeatedly ...
● Replace the DCS sensor.
sensor on board DX83.
sensor on board DX83.
sensor on board DX83.
not
occur with other causal
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
see
see
see
see
S. [ → 378]
see
S. [ → 378]
see
S. [ → 378]
64 95 258 D3632
D3632.076.01.03.02 09.2016