5 Messages
5.4 List of error messages
Error code
Description
E7 15 03
Wrong software constellation of
modules.
Error code
Description
E6 15 04
Product activation keys invalid or
not available.
Occurs after replacement of tube
assembly (DX6) or board DX11 and
possibly after software updates.
See also chapter "Measures
following replacement of boards
[ → 398]".
Error code
Description
E6 15 05
Unit serial number invalid or not
available.
Occurs during first power-on after
replacement of board DX6 or DX11.
See also chapter "Measures
following replacement of boards".
Error code
Description
E6 15 06
Wrong or invalid manufacturer code
on an module was detected.
Error code
Description
E6 15 07
Wrong or invalid manufacturer code
on an module was detected.
Error code
Description
E6 15 10
Update file for module is corrupt.
112
5.4.8

Location 15: Configuration, update

Actions required
E7 15 03
● Check the software versions of the unit (on
the info screen or by running service
routing S008.2), and run or repeat
software update or downgrade if
necessary.
Actions required
E6 15 04
● Entering the serial number
● Enter release key.
*) OI = Operating instructions
Actions required
E6 15 05
● Run service routine S008.3 and confirm or
enter the unit serial number on the unit.
Actions required
E6 15 06
● Replace the X-ray tube assembly.
Actions required
E6 15 07
● Replace board DX11.
Actions required
E6 15 10
● Obtain latest update file from the Sirona
Customer Service Center (CSC) or the
Sirona home page and perform software
update.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
see
S. [ → 267],
S. [ → 68]
see
see OI*
see
S. [ → 267]
see
S. [ → 372]
see
S. [ → 397]
see
S. [ → 68]
64 95 258 D3632