S020
S020.1
300
8.1.16

S020: Service for temple support

SA**
Function
Service for temple support
Yes
Programming the switching thresholds of the temple support
* SR=service routine, ** SA=security access
8.1.16.1

S020: Test step 1

Programming the switching thresholds of the temple support
This service routine can be used to change the factory-set switching
points of the temple width measurement used for setting the jaw width
("small", "normal" or "large") within certain limits.
The following drawing shows the motor position values (initial and
individual offset ranges) as well as the mechanical end stops of the
temple support.
The switch-over point between large ↔ normal is referred to as the lower
switch-over point (marked L = Lower) and the switch-over point between
normal ↔ small as the upper switch-over point (marked U = Upper).
These switch-over points can be adjusted with service routine S020.1.
Selection field
1
2
* Factory setting
1. Call service routine S020.1 [ → 249].
Parameters
L
U
After selecting the service routine, the lower switching point in
selection field 1 is displayed in the form "L: 0". The upper
switching point is displayed in selection field 2 in the form "U: 0".
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Range of values
-10...0*...+5
-5...0*...+13
64 95 258 D3632
D3632.076.01.03.02 09.2016