S034: Service for the digital cephalometer

SA**
Function
Service for the digital cephalometer
No
Display/calibrate center position of Ceph scan sensor axis
No
Display/calibrate center position of Ceph scan secondary diaphragm
axis
No
Moving to the Ceph packing position
No
User-specific determination of the shadowing limit for C1 and C2
No
User-specific determination of the shadowing limit for C3 and C3
30x23
* SR=service routine, ** SA=security access
8.1.20.1

S034: Test step 4

Calibrating / displaying he center position of ceph scan sensor axis
Tip: Move the unit to a comfortable working height with the Up/Down keys
on the control panel
When the service routine is called, it is not possible to adjust the height of
the unit.
1. Use UP/DOWN keys (A) on the control panel to move the unit up and
down and use patient symbol keys (F) to check the switching states.
2. Detach the cover to the cephalometer [ → 37].
3. Call service routine S034.4 [ → 249].
4. Touch the T key (A).
8.1 Perform service routines via the control panel
before
commencing calibration of the sensor axis.
After selecting the service routine, the current offset value for the
center position of the ceph scan
selection field 1.
The test mode is activated.
8 Service routines
sensor axis
is displayed in
309