10 Maintenance
10.11 Test exposures/Test images
A
438

Ceph - quality test exposure (2D)

10.11.2.7
You can use the
exposure with full rotation. This function enables you to simulate a quality
test, for example, in a similar way to an acceptance/constancy test.
No needle phantom is inserted into the bite block of the unit.
The Ceph sensor is inserted in the sensor holder of the
cephalometer.
1. Open the ear plug holders completely.
2. Turn the ear plug holders so that one ear plug is directly in front of the
image receptor.
3. Fold up the nose support.
4. Fit the contrast element (A) with its hole onto the ear plug.
NOTICE
The aluminum plate of the contrast element must face the X-ray tube
assembly.
"Ceph - quality control exposure"
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
to create a Ceph
64 95 258 D3632
D3632.076.01.03.02 09.2016